Boonton 4500B Peak Power Meter Manuale Utente Pagina 261

  • Scaricare
  • Aggiungi ai miei manuali
  • Stampa
  • Pagina
    / 372
  • Indice
  • SEGNALIBRI
  • Valutato. / 5. Basato su recensioni clienti
Vedere la pagina 260
Boonton 4500B RF Peak Power Analyzer
Remote Operation
5-61
READ:INTERval:MAXimum?
Description: For the specified channel, return the maximum instantaneous power or voltage between marker 1
and marker 2. The units will be the same as the specified channel.
Syntax: READ[1|2|3|4|6|7]:INTERval:MAXimum?
Returns: CC, maximum instantaneous power or voltage between markers
Where CC is the measurement condition code.
Valid Modes: Pulse and Modulated
READ:INTERval:MINimum?
Description: For the specified channel, return the minimum instantaneous power or voltage between marker 1
and marker 2. The units will be the same as the specified channel.
Syntax: READ[1|2|3|4|6|7]:INTERval:MINimum?
Returns: CC, minimum instantaneous power or voltage between markers
Where CC is the measurement condition code.
Valid Modes: Pulse and Modulated
READ:INTERval:MAXFilt?
Description: For the specified channel, return the maximum filtered power or voltage between marker 1 and
marker 2. The units will be the same as the specified channel.
Syntax: READ[1|2|3|4|6|7]:INTERval:MAXFilt?
Returns: CC, maximum filtered power or voltage between markers
Where CC is the measurement condition code.
Valid Modes: Pulse and Modulated
READ:INTERval:MINFilt?
Description: For the specified channel, return the minimum filtered power or voltage between marker 1 and
marker 2. The units will be the same as the specified channel.
Syntax: READ[1|2|3|4|6|7]:INTERval:MINFilt?
Returns: CC, minimum filtered power or voltage between markers
Where CC is the measurement condition code.
Valid Modes: Pulse and Modulated
Vedere la pagina 260
1 2 ... 256 257 258 259 260 261 262 263 264 265 266 ... 371 372

Commenti su questo manuale

Nessun commento