Boonton 4500B RF Peak Power Analyzer
Remote Operation
5-14
CALCulate:UNIT
Description: Set or return units for the selected channel. For power sensors, voltage is calculated with
reference to the sensor input impedance. Note that for ratiometric results, logarithmic units will
always return dBr (dB relative) and all other units return percent. Units for external trigger inputs
are always volts.
Syntax: CALCulate[1|2|5|6|7]:UNIT <asc>
Argument: <asc> = {dBm, W, V, dBV, dBmV, dBuV}