Read RSSI on PIM31; value should be +/-1 dB of to Signal Generator’s
Power level
Repeat the process you are no longer able to decrease the signal
frequency F2
.
Set PIM 31 Signal frequencies F1 and F2 back to Start and End
frequencies of PIM31 band.
Set Signal Generator to Frequency that results for IM3 product of
frequencies F1 and F2
Read the RSSI value on PIM31; the value should be equal to +/-1 dB of
the Signal Generator’s Power level
.
Change the power level of the Signal Generator by 5dB.
Never exceed -60dBm
Read RSSI on PIM31; value should be equal to +/-1 dB of the Signal
Generator’s Power level
.
Repeat until power level range of -60dBm to -120dBm has been covered
Switch RF Signal of Signal Generator OFF
If the readings from the previously listed procedures deviate by more than
+/- 1 dB, contact your nearest Boonton Service Center
.