Boonton 4540 Series RF Power Meter
Remote Operation
5-65
STATus:OPERation:NTRansition
Description: Set or return the value of the negative transition filter bitmask for the Operation status group.
Setting a bit in the positive transition filter causes a 1 to 0 (negative) transition in the
corresponding bit of the Operation Condition register to cause a 1 to be written in the associated
bit of the Operation Event register.
Syntax: STATus:OPERation:NTRansition <numeric_value>
Argument: <numeric_value> = 0 to 32767
Returns: 16-bit register value (0 to 32767)
Valid Modes: Any
STATus:OPERation:PTRansition
Description: Set or return the value of the positive transition filter bitmask for the Operation status group.
Setting a bit in the positive transition filter causes a 0 to 1 (positive) transition in the
corresponding bit of the Operation Condition register to cause a 1 to be written in the associated
bit of the Operation Event register.
Syntax: STATus:OPERation:PTRansition <numeric_value>
Argument: <numeric_value> = 0 to 32767
Returns: 16-bit register value (0 to 32767)
Valid Modes: Any
STATus:PRESet
Description: Sets SCPI enable registers and transition filters to the default state. The Operational Enable and
Questionable Enable mask registers are both cleared so an SRQ will not be generated for these
conditions. All bits for the device and questionable calibration registers are enabled. All positive
transition filters are enabled and all negative transition filters are cleared.
Syntax: STATus:PRESet
Argument: None
Valid Modes: Any